Instruction Manual
IM-106-5500, Original Issue
August 2005
CCO 5500
3-12
NOTE
As this value controls the calibration of the instrument do not change it without 
due consideration.
f. Plant Status Input 
Used to determine whether the plant is 
operating under correct conditions. 
There is a choice of three controls for 
plant status, only one can be used to 
control plant status at any one time; logic input, serial input and 
multiple input. The multiple input has five sub options; temperature 
threshold, oxygen threshold, water vapor threshold, detector level 
threshold and logic input. Any or all of these five options (in the 
multiple option) may be selected to determine plant status. 
g. Logic Input 
If the PS1 and PS2 terminals are linked in the signal processor the 
logic contact is made and the plant status is OFF. These terminals 
may be linked manually during a plant shut down, or they may be 
wired to a switch/contact outside the unit (e.g. a valve that opens and 
closes the duct). Press ENTER when the logic screen is displayed if 
this option is required. 
h. Serial Input 
If this option is selected the criteria controlling plant status are 
transmitted via the serial data link. Press ENTER on viewing the 
serial screen if this option is required to determine plant status. 
i. Multiple 
Press ENTER when the multiple display is seen and scroll through 
the following options. Select YES or NO for each option depending 
on whether it is to be used to determine plant status. If YES is 
chosen the configuration screen for that option will be shown. Enter 
the required value and press ENTER. The next option will then be 
shown, similarly select YES or NO and so on. After the final option 
(Logic Input) has been configured the display will revert to 'Multiple'. 
Scroll down using the arrow key and when EXIT is viewed press 
ENTER. The plant status is thus configured. 
j. Temperature Threshold 
A value is set here for the temperature 
threshold. While the temperature 
(taken from the normalizing 
temperature) is above the threshold 
value, plant status is ON. If the temperature drops below the 
threshold, plant status is OFF and only the seconds averaging stack 
will update. 
5 PARAMETERS
Plant Status I/P
5 Plant Status I/P
T th’hold
262
o