Table 6-2.  Configuration 
Problems
No display on signal processor
Possible Cause
Test
Result
Action
Mains input failure
Check 110/220V 
selection switch in 
power supply
Setting correct
Proceed to next test
Setting incorrect
Change setting & 
proceed to next test
Check power indication 
LED in power supply
LED illuminated
Mains OK - proceed to 
next possible cause
LED not 
illuminated
Proceed to next test
Check fuse in power 
supply
Fuse OK
Proceed to next 
possible cause
Fuse blown
Replace fuse
Power supply failure
Check power rail LEDs 
in signal processor
All LEDs 
illuminated
Supplies OK - 
proceed to next 
possible cause
LEDs not 
illuminated
Proceed to next test
Check fuse in signal 
processor
Fuse OK
Proceed to next 
possible cause
Fuse blown
Replace fuse
Connection Problem
Check wiring between 
signal processor & 
power supply
Wiring OK
Proceed to next test
Incorrect wiring
Correct wiring
Check ribbon cable 
connections in signal 
processor
Connections OK
Proceed to next 
possible cause
Connections 
loose
Secure connections
Signal processor 
failure
-
-
Contact your 
Rosemount Analytical 
supplier
Display nonsense on signal processor
Possible Cause
Test
Result
Action
Micro-processor fault
Reset by interrupting 
mains supply
Fault clears
No further action
Fault continues
Proceed to next 
possible cause
Program 
corruption/faulty micro 
PCB
-
-
Contact your 
Rosemount Analytical 
supplier