Emerson IM-106-5500 Manual Do Utilizador
Instruction Manual
IM-106-5500, Original Issue
August 2005
August 2005
6-3
CCO 5500
Table 6-2. Configuration
Problems
No display on signal processor
Possible Cause
Possible Cause
Test
Result
Action
Mains input failure
Check 110/220V
selection switch in
power supply
selection switch in
power supply
Setting correct
Proceed to next test
Setting incorrect
Change setting &
proceed to next test
proceed to next test
Check power indication
LED in power supply
LED in power supply
LED illuminated
Mains OK - proceed to
next possible cause
next possible cause
LED not
illuminated
illuminated
Proceed to next test
Check fuse in power
supply
supply
Fuse OK
Proceed to next
possible cause
possible cause
Fuse blown
Replace fuse
Power supply failure
Check power rail LEDs
in signal processor
in signal processor
All LEDs
illuminated
illuminated
Supplies OK -
proceed to next
possible cause
proceed to next
possible cause
LEDs not
illuminated
illuminated
Proceed to next test
Check fuse in signal
processor
processor
Fuse OK
Proceed to next
possible cause
possible cause
Fuse blown
Replace fuse
Connection Problem
Check wiring between
signal processor &
power supply
signal processor &
power supply
Wiring OK
Proceed to next test
Incorrect wiring
Correct wiring
Check ribbon cable
connections in signal
processor
connections in signal
processor
Connections OK
Proceed to next
possible cause
possible cause
Connections
loose
loose
Secure connections
Signal processor
failure
failure
-
-
Contact your
Rosemount Analytical
supplier
Rosemount Analytical
supplier
Display nonsense on signal processor
Possible Cause
Possible Cause
Test
Result
Action
Micro-processor fault
Reset by interrupting
mains supply
mains supply
Fault clears
No further action
Fault continues
Proceed to next
possible cause
possible cause
Program
corruption/faulty micro
PCB
corruption/faulty micro
PCB
-
-
Contact your
Rosemount Analytical
supplier
Rosemount Analytical
supplier