Table 6-3. Operational
Problems
Output reading permanently zero or full scale
Possible Cause
Possible Cause
Test
Result
Action
Incorrect normalizing
parameter setting
parameter setting
Enter parameter mode
and ensure all
parameters are as
required
and ensure all
parameters are as
required
Parameters
correct
correct
Proceed to next
possible cause
possible cause
Parameters
incorrect
incorrect
Enter correct
parameters
parameters
Incorrect calibration
(Set Cal value) zero
gas level
(Set Cal value) zero
gas level
Enter diagnostic mode
and observe the Y (60)
figure
and observe the Y (60)
figure
1800 <Y <2010
Instrument OK - true
level of gas in duct
(output permanently
zero)
level of gas in duct
(output permanently
zero)
Y <1800
Proceed to next
possible cause
possible cause
Detector levels low or
saturated
saturated
Enter diagnostic mode
and check the detector
levels and saturation
count
and check the detector
levels and saturation
count
Detector levels
OK
OK
Proceed to next
possible cause
possible cause
Detector levels
out of tolerance
out of tolerance
Reconfigure the
instrument and
consult the
fault-finding table
again
instrument and
consult the
fault-finding table
again
Output circuitry failure
Enter setup mode
(configure output and
attempt to set zero and
span
(configure output and
attempt to set zero and
span
Output responds
correctly
correctly
Reconfigure the
instrument and
consult the
fault-finding table
again
instrument and
consult the
fault-finding table
again
Output does not
respond correctly
respond correctly
Replace signal
processor
processor
Averages not updated
Possible Cause
Possible Cause
Test
Result
Action
Fault condition exists
Enter diagnostic mode
and check fault
condition
and check fault
condition
Fault condition
exists
exists
No fault condition
Proceed to next
possible cause
possible cause
Plant status contact in
signal processor has
been made
signal processor has
been made
Check plant status
input LED in signal
processor
input LED in signal
processor
LED off
Operation correct -
stable gas level
stable gas level
LED on
Plant status in use -
instrument will not
update until plant on
instrument will not
update until plant on