Emerson IM-106-5500 Manual Do Utilizador

Página de 108
Instruction Manual
IM-106-5500, Original Issue
August 2005
6-7
CCO 5500
Table 6-3.  Operational 
Problems
Output reading permanently zero or full scale
Possible Cause
Test
Result
Action
Incorrect normalizing 
parameter setting
Enter parameter mode 
and ensure all 
parameters are as 
required
Parameters 
correct
Proceed to next 
possible cause
Parameters 
incorrect
Enter correct 
parameters
Incorrect calibration 
(Set Cal value) zero 
gas level
Enter diagnostic mode 
and observe the Y (60) 
figure
1800 <Y <2010
Instrument OK - true 
level of gas in duct 
(output permanently 
zero)
Y <1800
Proceed to next 
possible cause
Detector levels low or 
saturated
Enter diagnostic mode 
and check the detector 
levels and saturation 
count
Detector levels 
OK
Proceed to next 
possible cause
Detector levels 
out of tolerance
Reconfigure the 
instrument and 
consult the 
fault-finding table 
again
Output circuitry failure
Enter setup mode 
(configure output and 
attempt to set zero and 
span
Output responds 
correctly
Reconfigure the 
instrument and 
consult the 
fault-finding table 
again
Output does not 
respond correctly
Replace signal 
processor
Averages not updated
Possible Cause
Test
Result
Action
Fault condition exists
Enter diagnostic mode 
and check fault 
condition
Fault condition 
exists
No fault condition
Proceed to next 
possible cause
Plant status contact in 
signal processor has 
been made
Check plant status 
input LED in signal 
processor
LED off
Operation correct - 
stable gas level
LED on
Plant status in use - 
instrument will not 
update until plant on