PFD1/PFS1 – OPERATING INSTRUCTIONS 
 
INDEX 
DATE 
N° 
03 
27/09/01 
DC01020US 
 
3/11 
 
II.2  – Reception 
 
Upon receipt of the pump, please check that: 
• 
The carton has not been damaged in transit. If there is any visible damage, immediately contact the 
carrier.  
• 
The pump is not damaged. If there are signs of damage, you should report this immediately to 
SGPPL ASTI or your local distributor. 
• 
An operating instruction manual has been included in each package. Please request another copy if 
it has not been included. 
 
 
III – Installation and Operation 
 
III.1  – Testing 
 
All pumps are tested with DI water at the factory in our clean room for: 
• 
Maximum flow rate with no back pressure 
• 
Minimum flow rate with no back pressure 
• 
Flow rate with 58 PSI (4 bar) discharge pressure 
• 
Checked for leakage 
 
III.2  – Set up 
 
The pump must be installed  horizontally as shown on general arrangement drawing (see appendix 
“APP 1 EXT”). This drawing also shows the overall dimensions of the pump etc. 
The pump must be  positioned on its feet. If not, the check valves will not seat correctly and the pump 
may malfunction. 
 
III.3  – Connections 
 
III.3.1  – Air/Nitrogen connection 
 
The pump must be connected to a clean dry air or nitrogen supply. On no account should the air/nitrogen 
supply be lubricated, oil or water droplets will cause the shuttle valve to malfunction. 
Minimum and maximum supply pressure must be between 29 and 72.5 PSI (2 and 5 bar).  
For optimum pump operation, we recommend a supply pressure of 51 PSI (3.5 bar). 
The ID of the tube supplying the dry air/nitrogen should not exceed 3/32” (2.5 mm). The tube length 
between the pump and on/off valve should be between 10 feet minimum (3 m) and 20 feet maximum 
(6 m). 
When in aggressive conditions (acid vapors), it is advised to canalize outlet with a tube of minimum  
ID 1/4” (6 mm). 
The pneumatic on/off valve must be 3-way to ensure the shuttle valve on the pump resets itself when the 
pump is switched off. The flow control valve must be positioned before the 3-way on/off valve  
(see appendix “APP 1 CAB”). 
 
A remote control box with on/off switch and needle valve (P/N 21 000 04) is available as an optional 
extra.