Canon GC-251 Manual Do Utilizador

Página de 88
 
CANON    Digital Galvano Scanner System
 
KP-1SM Series + GC-251 
Users Manual 1.00 
 
 
 
  
6
2. Specifications 
 
2.1. 
Galvano Motor (KP-1SM30, KP-1SM100) 
Performance and shape 
 KP-1SM30 
KP-1SM100 
Conforming 
beam diameter 
Φ10 to φ20 
Φ20 to φ30 
Scan angle 
±20 deg 
±20 deg 
Number of encoder   
pulses 
83328 pulses/rotation 
83328 pulses/rotation 
Command resolution 
0.036 μrad 0.036 
μrad 
Torque constant 
0.041 Nm/A 
0.093 Nm/A 
Weight 0.95kg 
1.65kg 
    (Reference) Command resolution 
                The controller of this system divides one encoder cycle into 2048. 
        KP-1SM30,SM100 is controlled by the following command resolution: 
               83328 pulses × 2048 divisions = 170655744 pulses/rotation (360°) 
360°/ 170655744 = 0.000002109° = 0.036urad  
 
Environmental conditions 
Operating temperature and humidity 
23℃±10℃, 80% RH or less (No condensation)
Storage temperature and humidity 
–20 to 60ºC, 90% RH or less (No condensation)
Note: The above operating temperature and humidity conditions depend on the operating and heat 
radiation conditions.