Oritron ICS-2100 Manual Do Utilizador
ICS-2100 Ion Chromatography System
198
Doc. 065291-01 3/09
A.6
Eluent Generation
Eluent Types
KOH, LiOH, NaOH, K
2
CO
3
, K
2
CO
3
/KHCO
3
, MSA
Gradient Profiles
Optional feature; combination of unlimited number of linear, convex,
and concave positive and negative gradient profiles
and concave positive and negative gradient profiles
Concentration
Increments
0.01 mM
Flow Rate
0.1 to 3.0 mL/min
Maximum
Operating
Pressure
21 MPa (3000 psi)
Maximum
Solvent
Concentration
•
Anion: 25% methanol
•
Cation: No solvents
Concentration
Range
Dependent on the EluGen cartridge type, the cartridge configuration
(single or linked dual), and the flow rate. Refer to
(single or linked dual), and the flow rate. Refer to
and
.
EluGen Cartridge
Eluent Concentration Range
K
2
CO
3
0.1 to 15 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow in mL/min
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow in mL/min
KOH
0.1 to 100 mM at
0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
where X = 100/flow in mL/min
LiOH
0.1 to 80 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 80/flow in mL/min
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 80/flow in mL/min
MSA
0.1 to 100 mM at
0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
where X = 100/flow in mL/min
NaOH
0.1 to 100 mM at
0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow mL/min
where X = 100/flow mL/min
Table A-6. Eluent Concentration Ranges for Single-Cartridge Configurations