Emerson 2400S Manual Do Utilizador

Página de 50
14
Micro Motion
®
 Model 2400S Transmitters 
I/O Wiring – Model 2400S Analog Transmitters
Figure 3-2
HART/analog single-loop wiring – Internal power 
Figure 3-3
Basic mA output wiring – External power 
HART-compatible 
host or controller
For analog communications:
• 820 Ohm maximum loop 
resistance
For HART communications:
• 600 Ohm maximum loop 
resistance
• 250 Ohm minimum loop 
resistance
+
Note: See Figure 3-5 for voltage and 
resistance values.
VDC
R
load
+
+
CAUTION
Excessive current will damage the transmitter.
Do not exceed 36 VDC input. Terminal current must be less than 500 mA.
Note: Channel B (terminals 3 and 4) 
can also be configured this way.