5.4 Adjustment after replacement of unit
After completion of the assembly work mentioned above, use the following procedures for adjust-
ment and setting. Adjustment should be performed using the HHC.
(1) After replacement of electronics unit  (including replacement of internal parts)
(
2) After replacement of detecting unit (including replacement of internal parts)
Adjustment
S
tep
Adjustment item
Relevant page
1
Constant current output (output circuit)
P19
2
TAG.  No.
P11
3
Type
P12
4
Industrial value unit
P13
5
Range (zero/span)
P14
6
Zero/span adjustment
P18
7
Damping
P15
8
Output mode (LIN/
 selection)
P16
9
Burnout direction
P17
10
Lock of adjustment function
P22
Step
Adjustment item
Relevant page
1
Range (zero/span)
P14