Hoffman Media EDD-4C Manual Do Utilizador

Página de 29
  23
Section 5: Maintenance and Troubleshooting 
 
Table 4: Routine Maintenance 
Task 
Frequency 
Check for oil leaks 
Monthly 
Oil filter change (the oil filter used can be an 
automotive fuel filter with ¼" fittings) 
Every six months 
Check for normal operation 
•  The displayed drip rate is equal to the Normal (pump 
on) mode drip rate setting, within a -1, +2 digit 
tolerance  
Weekly 
 
Table 5: Troubleshooting 
Problem 
Possible Cause 
Solution 
The display and LEDs are not lit. 
•  The unit is not powered. 
•  Check power in the connection 
box. 
An unusual result appears on the 
display 
•  Changing the mode toggle 
switch (item 8 in Figure 6 on 
page 13). 
•  Ignore. 
You want to quickly stabilize the 
Dripmaster for a new drip rate 
•  You just refilled the oil tank. 
 
 
 
•  You just increased the drip rate 
setting in the Normal (pump on) 
mode. 
•  With the Mode toggle switch 
(item 2 in Figure 6 on page 13) 
in automatic (AUTO), hold 
down the Close push button 
momentarily to slow down the 
drip rate. 
•  With the Mode toggle switch in 
position AUTO, hold down the 
Open push button momentarily 
to speed up the drip rate. 
A suspected fault has occurred in 
the system; the drip rate is much 
higher or lower than expected. 
•  Circuit failure 
•  Set the Mode switch to MAN
and set the drip rate manually 
using the Open and Close 
push buttons. See Figure 6.