SECTION 2—OPERATING THE BATH LIFT
11
15. Lock the two rear locking casters on the lift by pressing down on the locking pedal 
with your foot. Refer to FIGURE 2.1.
16. Lock the bathing unit door. Refer to the bathing unit’s owner’s manual.
Raising/Lowering the Bath Lift
NOTE: For this procedure, refer to 
FIGURE 2.3.
To raise the bath lift, press the up arrow 
button on the hand pendant.
To lower the bath lift, press the down 
arrow button on the hand pendant.
Using the Scale (Model 1900S ONLY)
NOTE: For this procedure, refer to FIGURE 2.4.
1. With the seat empty and arms lowered, turn on the scale by touching the “on” button 
on the face of the scale.
2. Press the “ZERO” button on the face of the scale. 
NOTE: The digital readout shows 0.0.
3. Using the “UNITS” button, select either kilograms or pounds.
4. Transfer the resident on the seat of the lift and lower the arms.
5. Raise the resident. 
NOTE: It is only necessary to raise the resident 
until his/her feet are off the floor and not 
touching the legs of the lift.
6. Press the “LOCK” key
7. Read and record the weight of the resident 
on the readout of the scale. 
Emergency Stop
NOTE: For this procedure, refer to 
FIGURE 2.5.
• Press the RED emergency stop button 
on the control box to stop the lift seat 
and resident from raising or lowering.
• To disengage, rotate the RED 
emergency stop button CLOCKWISE 
until it pops out.
RED Emergency Stop Button
Control Box
Press to Stop 
Lift Seat
Rotate CLOCKWISE to 
Disengage Emergency Stop