SECTION 3—MAINTENANCE AND TROUBLESHOOTING
23
Troubleshooting
NOTE: If problems are not remedied by the suggested means, please contact your dealer or 
Invacare.
SYMPTOMS
FAULTS
SOLUTION
Seat and/or upper
column feels loose.
Nylon bushing is loose.
Casters/brakes noisy or 
stiff.
Fluff or debris in bearings.
Clean the or replace the casters.
Electric actuator fails to 
lift when button is 
pressed.
Hand-control or actuator con-
nector loose.
Batteries low.
RED emergency stop button 
pressed IN.
Battery not connected prop-
erly to control box.
The connecting terminals are 
damaged.
Electric actuator in need of ser-
vice or load is too high.
Check connections.
Rotate RED emergency stop button CLOCK-
WISE until it pops out. 
Contact your Dealer.
Controller beeps.
Low battery power.
Charge battery. 
Unusual noise from 
actuator.
Actuator is worn or damaged 
or spindle is bent.
Contact Invacare.
Lift arms keep falling.
Jam nuts are loose.
Scale does not work 
properly.
Battery failure
Check battery. Replace if necessary
Battery has been 
replaced and scale still 
does not work properly
Contact Invacare.