Main menu 
(continued)
„
„
Equipment required for calibration
• Konica Minolta CA-310 display color analyzer (optional accessory)
• Konica Minolta CA-PU12 or CA-PU15 standard measurement probe (optional accessory)
• A RS-232C cable (male to male, straight) (optional accessory)
Connect the RS-232C input terminal  on this unit to the RS-232C terminal  on the display color analyzer using a straight cable.
• Set the Baud rate of the display color analyzer to 38,400.
This unit
Konica Minolta CA-PU15
standard measurement probe
(optional accessory)
RS-232C cable
(optional accessory)
Konica Minolta CA-310 display
color analyzer (optional accessory)
RS-232C input 
terminal (9P) 
Analyzer Side
(Straight)
On the monitor
Pin number Signal
Pin number Signal
1
CD
1
N.C.
2
RXD
2
TXD
3
TXD
3
RXD
4
DTR
4
DSR
5
GND
5
GND
6
DSR
6
DTR
7
RTS
7
CTS
8
CTS
8
RTS
9
GND
9
N.C.
• Turn on this unit and perform adequate aging (approx. 1 hour) before starting calibration.
• Make the room dark so that no external light can enter the standard measurement probe before starting the calibration. If external light 
enters the probe, the low brightness characteristics may not be calibrated correctly.
• LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small differences in values after 
calibration.
To further fine tune the monitor, set GAIN and BIAS for R, G and B in the VAR mode of COLOR TEMP.
• Do not apply the probe to a WFM/VECTOR display on the screen.
• If you are using CA-210, consult the supplier of the unit.
„
„
CALIBRATION
Select “CALIBRATION” in the “SYSTEM CONFIG” menu to open the following menus.
Sub menu
Settings
Description
AUTO CALIBRATION 
*
8
Connect a CA-310 display color analyzer and use this submenu to make a 
calibration.
Select “AUTO CALIBRATION” and select “YES” in the confirmation screen that 
appears to start calibration.
RESET 
*
9
Returns calibration data to their factory defaults.
Select “RESET” and select “YES” in the confirmation screen that appears to return 
calibrated values to their factory defaults.
*
8
 
“EXECUTING” is displayed during “AUTO CALIBRATION” and “COMPLETE” appears when calibration ends. “INCOMPLETE” ap-
pears if calibration could not be completed.
*
9
 
When “RESET” ends, “COMPLETE” appears.
Performing AUTO CALIBRATION
1. 
Turn the rotary knob in the “SYSTEM CONFIG” menu to 
select the “CALIBRATION” submenu and press the rotary 
knob.
2. 
Turn  the  rotary  knob  to  select  the  AUTO  CALIBRATION 
submenu and press the rotary knob.
MENU EXIT
RETURN
RETURN
SEL.
ENTER
       [AUTO CALIBRATION]
               YES
               NO
3. 
Select “YES” and press the rotary knob.
This starts calibration.
MENU EXIT
RETURN
RETURN
SEL.
ENTER
       [AUTO CALIBRATION]
            xxxxxxxxx
xxxxxxxxx indicates that one of the following messages with 
the meaning listed below appears.
EXECUTING:  Operation in progress
COMPLETE: 
Operation completed
INCOMPLETE:   Operation incompleted.
 
 
 
 
 (Check the RS-232C connection or display 
color analyzer connection.)
4. 
Turn the power off and then on after performing auto cali-
bration.