5.2  Internal AM Mode 
Introduction 
   The Internal AM mode generates an amplitude-modulated signal form with fixed carrier and modulating 
frequencies.  An internally generated sinusoid is used as a modulating signal to vary the amplitude of a carrier 
sinusoid. The modulation waveform is not suppressed carrier; i.e. a fixed amount of carrier power is always present 
in the modulated signal. . Internal AM mode could be entered by pressing “Mode”> “AM” >”1” key at any time.  
Internal AM Mode Parameters  
   The Internal AM mode has the following front panel display:  
 
Figure 5.2-1: Internal AM mode display  
1. Modulating Frequency  
   In this field enter the frequency of the modulating sinusoid, from 0 Hz to 10,000 Hz in 1 Hz steps.  
2. Percentage modulation  
     
In this field enter the degree to which the modulating signal is allowed to change the carrier amplitude. You may 
enter from 0% (no change) to 100% (maximum change) in 1% steps. 
   To enter a value, type 1 to 3 numeric digits. If you type 3 digits, the value is automatically entered.  If you only 
type 1 or 2 digits, you can press the MHz key or the KHz key or the Hz key (all are equivalent) to enter the 
percentage value.  
3. Carrier Frequency  
   In this field enter the frequency of the carrier, from 0 Hz (DC) to 21.5 MHz in 0.01 Hz steps.  
4. PEP Level  
     
In this field enter the output level, from 4 mVp -p to 10 Vp -p in 1 mV steps or from -44.0 dBm to +24.0 dBm in .1 
dBm steps.  
   NOTE: The level specified in Internal AM mode is the Peak Envelope Power, or PEP. As a peak-to-peak value, 
this level represents the maximum peak-to-peak voltage swing that will result from a 100 percent modulated carrier. 
If no modulating signal were applied, the unmodulated carrier would have only half the peak-to-peak swing of a 
100% modulated carrier.  
   NOTE: The level specified is a 50O loaded level. This  is the level of the signal, which will appear across a 50O 
load connected to the SIG Out connector. Into an open circuit, the output swing will be twice the value entered.  
Offset  
  You can enter an offset voltage for the output signal. For more information on output offsets refer to section 4.4.  
Ext Gating Input  
   This TTL compatible input can be used to gate the output signal on or off.  A logic high level turns off the output.  
For further information on the Ext Gating Input, refer to section 2.0. 
 Int AM:  1,000 Hz 
             Percent Mod: 100% 
 1,000,000.00 Hz  
 
PEP level:  -10.0 dBm 