B&K 4070A Manual Do Utilizador

Página de 88
 
 
BK Precision 4070A User Manual Rev.2.2 
25 
5.12  Internal SSB Mode 
Introduction 
   Internal SSB mode generates a Single Sideband modulated signal of fixed carrier frequency. Internally generated 
signal is used to modulate either Upper sideband or Lower sideband of a carrier signal. Modulating, carrier 
frequencies and output level with DC offset values could be manually entered on the keypad. External signal gating 
option is available in this mode. Internal SSB mode could be entered by pressing “Mode”> “SSB” >”1” key at any 
time. 
Internal SSB Mode Parameters  
   The Internal SSB mode has the following front panel display:    
Figure 5.12-1: Internal SSB mode display 
1. Modulating Frequency  
   In this field enter the frequency of the modulating sinusoid.  You may enter from 0 Hz to 1 MHz in 1 Hz steps.  
   NOTE: It is possible to enter values for the modulating and carrier frequencies such that the upper sideband can 
exceed 21.5 MHz or the lower sideband can go below 0 Hz.  Care should be taken not to do this since distortion of 
the output waveform will result.       
2. Upper/Lower Sideband selections   
   This parameter selects which sideband will be generated.  Single sideband modulation specifies that all energy in 
the modulated waveform should be either above the carrier frequency or below it.  Upper sideband places all signal 
energy above the carrier frequency, and Lower sideband places all signal energy below the carrier frequency. To 
select Lower sideband, press 0.  To select Upper sideband, press 1. Pressing any arrow key or rotating the wheel will 
toggle the sideband selection between upper and lower.  
3. Carrier Frequency  
   In this field enter the frequency of the carrier.  You may enter from 0 Hz (DC) to 21.5 MHz in 0.01 Hz steps.  
   NOTE: It is possible to enter values for the modulating and carrier frequencies such that the upper sideband can 
exceed 21.5 MHz or the lower sideband can go below 0 Hz.  Care should be taken not to do this since distortion of 
the output waveform may result.  
4. Level  
   In this field enter the output level, from 4 mVp -p to 10 Vp -p in 1 mV steps or from -44.0 dBm to +24.0 dBm in 
0.1 dBm steps. 
  NOTE: The level specified is a 50O loaded level. This is the level of the signal, which will appear across a 50O 
load connected to the SIG Out connector. Into an open circuit, the output swing will be twice the value entered. 
 
Offset  
   You can enter an offset voltage for the output signal. For more information on output offsets refer to section 4.4.  
Ext Gating Input  
   This TTL compatible input can be used to gate the output signal on or off.  A logic high voltage turns off the 
output. For further information on the Ext Gating Input, refer to section 2.0.  
Int SSB: 1,000 Hz 
 
         Upper Sideband 
1,000,000.00 Hz  
 
 
        -10.0 dBm 
1
3
2
4