7.4 Arbitrary Waveform Mode 
 
Figure 7.4-1: Arbitrary Waveform Mode display  
1. Int Clock / Ext Clock / Lock Master Lock slave mode  
In this field select the clocking mode for the arbitrary waveform system: 
  Internal Clock   - The sample clock is generated internally. The clock frequency is entered directly via the front 
panel. 
  External Clock - The sample clock (40 MHz max) is taken from the External Arb Clock input on the back panel.  
*** Warning *** 
TTL/CMOS levels (0V to +5V) only
  Lock Master  
- The unit provides the master timing signals for locking multiple 4070A Arb generators together. 
  Lock Slave  
- The unit receives all timing signals from a master unit. Used when locking multiple 4070A Arb’s 
together. For more information on locking Arb’s together, see the section 7.9. 
2. Phase 
   This field is used during Multiple Unit Locking and only appears when the unit is in Lock Slave mode. It allows 
you to specify a phase offset of the arbitrary waveform relative to the waveform being generated by the Master Unit. 
For more information, see section 7.9.
 
3. Continuous/Triggered mode  
   In this field select whether the waveform is generated continuously or on a single event basis.  If you select 
Continuous mode, then the waveform is immediately restarted once the last point is reached.  If you select a 
Triggered mode, then the waveform generation is halted after the last point is reached, and the waveform is not 
restarted again until another trigger occurs.  The trigger can come from three sources:  
 
 
        1. Pressing the 
Trigger
 key  
 
 
        2. Applying a low-to-high transition on the Ext. Trig In connector  
 
 
        3. Sending an ASCII "T" to the RS-232 port  
 
   The 4070A will simultaneously accept a trigger from all of the above sources.  
To set Continuous-triggering mode, press 1.   
To set Triggered mode, press 0.  
Pressing any arrow key or rotating the wheel will toggle the run mode between Continuous and Triggered. 
4. Clock Frequency 
   In this field enter the Arbitrary Waveform clock frequency. This is the frequency at which your waveform values 
are sent to the Digital to Analog converter (DAC). You may enter from 0 Hz (DC) to 40.0 MHz in 0.01 Hz steps.  
 If a value of 0 Hz is entered, the arbitrary waveform clock is halted. 
5. Level 
   In this field enter the output level, from 4 mVp -p to 10 Vp -p in 1 mV steps. This level is the peak-to-peak voltage 
swing of the waveform across a 50O loaded, into an open circuit, the voltage swing will be twice the specified peak-
to-peak value.
           
Offset     
   
Enter a DC offset voltage by pressing the Offset key.  For more information refer to section 4.4.
 
Arb Mode  Int Clock   
     Phase: 126.35 deg 
Cont     Clock:    1,000,000.00 Hz          10,000 mV 