Controller Operation
This chapter discusses in detail how to operate the dry-well temperature con-
troller using the front control panel. Using the front panel key-switches and
LED display the user may monitor the well temperature, set the temperature
set-point in degrees C or F, monitor the heater output power, adjust the control-
ler proportional band, and program the calibration parameters, operating pa-
rameters, and serial interface configuration. Operation of the functions and
parameters are shown in the flowchart in Figure on page 20. This chart may
be copied for reference.
In the following discussion a button with the word SET, EXIT, UP, or DOWN
indicates the panel button, while the dotted box indicates the display reading.
Explanations of the button or display reading are to the right of each button or
display value.
6.1
Well Temperature
The digital LED display on the front panel allows direct viewing of the actual
well temperature. This temperature value is normally shown on the display. The
units, C or F, of the temperature value are displayed at the right. For example,
100.00C
Well temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing and holding the “EXIT” button.
6.2
Temperature Set-point
The temperature set-point can be set to any value within the range and with res-
olution as given in the specifications. Be careful not to exceed the safe upper
temperature limit of any device inserted into the well.
Setting the temperature involves selecting the set-point memory and adjusting
the set-point value.
6.2.1
Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be
quickly recalled to conveniently set the calibrator to a previously programmed
temperature set-point.
To set the temperature, first select the set-point memory. This function is ac-
cessed from the temperature display function by pressing “SET”. The number
of the set-point memory currently being used is shown at the left on the display
followed by the current set-point value.
100.00C
Well temperature in degrees Celsius
19
6 Controller Operation
Well Temperature