changed states. This feature allows for minimal block temperature scanning
times and improved accuracy.
times and improved accuracy.
Example 2:
With the instrument displaying 25°C, set the set-point to 100°C and enable the
Scan Hold feature. When the thermal switch changes states at 50°C, the con-
troller automatically sets a new set-point of 50°C and the status flashes either
"o or c".
Scan Hold feature. When the thermal switch changes states at 50°C, the con-
troller automatically sets a new set-point of 50°C and the status flashes either
"o or c".
Note: There is a delay between the displayed temperature and the actual tem-
perature of the thermal switch. To achieve better accuracy when the thermal
switch is tripped, the scan rate should be set for a slower time. The slower the
scan rate the more time required to perform a test. The operator must determine
the optimal scan rate for the specific application.
perature of the thermal switch. To achieve better accuracy when the thermal
switch is tripped, the scan rate should be set for a slower time. The slower the
scan rate the more time required to perform a test. The operator must determine
the optimal scan rate for the specific application.
6.5
Ramp and Soak Program Menu
The ramp and soak program feature allows the user to program a number of
set-points and have the dry-well automatically cycle between the temperatures,
holding at each for a determined length of time. The user can select one of four
different cycle functions.
set-points and have the dry-well automatically cycle between the temperatures,
holding at each for a determined length of time. The user can select one of four
different cycle functions.
The program parameter menu is accessed by pressing “SET” and then “UP”.
20.00C
Well temperature
S+U
Access program menu
ProG
Program menu
Press “SET” to enter the program menu
S
Enter program menu
6.5.1
Number of Program Set-points
The first parameter in the program menu is the number of set-points to cycle
through. Up to 8 set-points can be used in a ramp and soak program.
through. Up to 8 set-points can be used in a ramp and soak program.
Pn=8
Number of program set-points
Use the “UP” or “DOWN” buttons to change the number from 2 to 8.
Pn=3
New number of program set-points
Press “SET” to continue. Pressing “EXIT” causes any changes made to the pa-
rameter to be ignored.
rameter to be ignored.
9011 Dual-Well Calibrator
User’s Guide
User’s Guide
24