9
5
7
3
8
4
6
1
2
750W
1500W
6V - 60V
80V - 600V
AC INPUT
ON
OFF
+V
-V
NOT ACTIVE
J2 SW1
GPIB
ANALOG PROGRAMMING
+S+LS NC -LC-S
1 2 3 4 5 6 7 8 9
10/100 Ethernet
LINK
TX
J1
!
!
1 – AC input connector
Wire clamp connector for 1500W output models.
IEC connector for 750W output models.
2 – DC output connector
Wire clamp connector for 80V to 600V models.
Bus bars for 6V to 60V models.
Bus bars for 6V to 60V models.
3 – USB connector
Connector for connecting to a USB interface. See chapter 4 for setup.
4 – LAN connector
Connector for connecting to a LAN interface. LINK LED indicates link integrity.
TX LED indicates LAN activity. See chapter 4 for LAN setup.
TX LED indicates LAN activity. See chapter 4 for LAN setup.
5 – Analog Programming
connector
connector
Connector for the analog interface. Includes output voltage and current limit
programming and monitoring signals, Shut-Off control (electrical signal),
Enable/Disable control (dry-contact), power supply ok (Power Supply OK) signal
and operation mode (CV/CC) signal. (See next page for details)
programming and monitoring signals, Shut-Off control (electrical signal),
Enable/Disable control (dry-contact), power supply ok (Power Supply OK) signal
and operation mode (CV/CC) signal. (See next page for details)
6 – SW1 setup switch
Nine-position switch for selecting remote programming and monitoring modes
for Output Voltage, Current Limit and other control functions. (See next page for
details)
for Output Voltage, Current Limit and other control functions. (See next page for
details)
7 – Remote Sense connector Connector for making remote sensing connections for regulating the load
voltage and compensating for wiring voltage drop. (See next page for details)
8 – GPIB connector
Connector for connecting to a GPIB interface. See chapter 4 for setup.
9 – Ground screw
M4x8 screws for making chassis ground connections
WARNING
SHOCK HAZARD The power cord provides a chassis ground through a third
conductor. Be certain that your power outlet is of the three-conductor type
with the correct pin connected to earth ground
conductor. Be certain that your power outlet is of the three-conductor type
with the correct pin connected to earth ground