Invacare Compressor Manual Do Utilizador

Página de 40
SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES
35
O
2
 BELOW NORMAL 
(YELLOW) light stays On
1. Ensure the concentrator has warmed up for at 
least 20 minutes.If it has not been 20 minutes, 
turn the compressor Off until the concentrator 
warms up. If the YELLOW light still remains on 
proceed to STEP 2.
2. Inspect the connection between the compres-
sor and the concentrator for damage or pinching. 
If damaged, replace. If pinched, straighten tubing. 
Turn the compressor Off for at least 10 
seconds. 
Turn the compressor On. If the YELLOW light 
still remains on after 10 minutes, proceed to 
STEP 3.
3. The concentrator may need service, contact 
your home health care provider or Invacare.
Fill times exceed 
maximum as 
recommended fill times 
OR GREEN 
filling light stays On.
1. Ensure the cylinder regulator is set to Off. 
2. Check the recommended fill times shown in 
3. Check that the cylinder is connected properly. 
Turn the compressor 
Off. Remove the cylinder and reinstall. Turn the 
compressor ON.
4. Check the compressor/concentrator line for 
proper connection and/or damage (i.e. - leaks or 
kinks).
5. Check the concentrator liter flow setting is at 
three L/min. or less. If necessary, adjust the liter 
flow setting (refer to the concentrator owner’s 
manual). If the GREEN filling light does not go Off, 
the compressor and/or concentrator may need 
service. Contact your home health care 
provider or Invacare.
Excessive noise when 
compressor is On
1. Call home health care provider or Invacare for 
service.
Hissing sound from Regu-
lator when flow selector 
is Off
1. Turn flow selector Off.
2. If hissing sounds continue, call home health 
care provider or Invacare.
TROUBLESHOOTING
PROBLEM
SOLUTION