SECTION 4—CLEANING AND DISINFECTING
30
4. Thoroughly clean the nebulizer and mouthpiece in hot 
water/detergent solution.
5. Remove from the water/detergent solution and rinse with 
clear hot tap water.
6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.
NOTE: If using medical disinfectant cleaners, follow 
manufacturer’s instructions carefully.
8. Keep the outer surface of the tubing dust-free by wiping 
regularly.
NOTE: The nebulizer air tubing does not have to be washed 
internally because only filtered air passes through the tubing
Changing the Nebulizer
The nebulizer is designed for single patient use.
The nebulizer must be replaced after a long period of 
inactivity, or when it is obstructed by dry medication, dust, 
etc. The reusable nebulizer should last from six months to one 
year with general use. The disposable nebulizer should last 30 
days.