Analog Devices UG-001 Manual Do Utilizador

Página de 24
UG-001 
Evaluation Board User Guide
 
Rev. 0 | Page 10 of 24 
Adjusting the Amplitude of the Input Signal 
Next, adjust the amplitude of the input signal for each channel 
as follows: 
1.  Adjust the amplitude of the input signal so that the 
fundamental is at the desired level. (Examine the Fund 
Power
 reading in the left panel of the VisualAnalog FFT 
window.) If the gain pin voltage is too low, it is not possible 
to reach full scale without distortion. Use a higher gain 
setting or a lower input level to avoid distortion. This also 
depends on the PGA gain setting, which can be 30 dB, 
27 dB, 24 dB, or 21dB. See Figure 15 and Figure 16
07
78
2-
02
6
 
Figure 15. VisualAnalog, Graph Window 
0
77
82
-02
7
 
Figure 16. VisualAnalog, Formatted FFT Plot 
2.  Repeat this procedure for the other seven channels.  
3.  Click the disk icon within the Graph window to save the 
performance plot. See Figure 17 for an example. 
0
–130
0
2
0
778
2-
1
19
FREQUENCY (MHz)
AM
P
L
IT
UD
E
 (
d
BF
S
)
5
–10
–20
–30
–40
–50
–60
–70
–80
–90
–100
–110
–120
5
10
15
20
f
IN
 = 3.5MHz @ –1dBFS
LNA = 6×
V
GAIN
 = 1V
FILTER TUNED
HPF = 700kHz
 
Figure 17. Typical FFT, AD9272/AD9273 
USING THE INTEGRATED CROSSPOINT SWITCH 
(CW DOPPLER MODE) 
To examine the spectrum of the CW Doppler integrated 
crosspoint switch output, use the following procedure: 
1.  Complete the steps in the Configuring the Board an
Using the Software for Testing sections to ensure that the 
board is set up correctly. 
2.  Optionally, remove the voltage source from the gain pin. It 
does not affect the CW Doppler output. 
3.  Connect the dc voltage source to P601, connecting the  
−5 V pin, the 0 V ground pin, and the +5 V pin as shown 
in Figure 1. These benchtop linear supplies should each have 
100 mA of current capability.  
4.  Place jumpers on the top pin pairs of P606 or P607 to 
connect CWD2+/CWD2− to CWD5+/CWD5− to the 
IOP/ION buses. This directs each of these connections to 
the output amplifier for display. 
Note that the CWD0±/CWD1±/CWD6±/CWD7± outputs 
are configured and biased to interface with the 
evaluation board. The AD9272/AD9273 is specially designed 
to snap onto the AD8339 evaluation board to allow the user to 
evaluate a larger portion of this common signal chain. For 
detailed instructions about enabling this function, send an 
email to 
5.  Use a 1 m, shielded, RG-58, 50 Ω coaxial cable to connect 
the spectrum analyzer to J402 (labeled AOUT on the 
evaluation board). 
6.  In the ADCBase 0 tab of the SPI Controller, find the 
MODES(8)
 box. Select the CW Mode option (see Figure 18).