unit may be damaged or fail prematurely. Such damage will not be 
repaired under warranty. 
 
Units should be installed in a clean, dry environment with an ambient 
temperature that is as stable as possible. Avoid areas with strong magnetic 
fields, strong air flows or excessive vibration. 
 
In order to operate the differential pressure across the controller should be 
in the range 15-45psid (1-3 bar). For optimum performance a differential 
pressure of 25psid is recommended. 
 
For Example, consider the following system:  
 
 
 
 
 
The differential pressure across the flow controller in this system would be 
100 psi – 14 psi = 86 psid. Consequently the flow controller would NOT be 
able to control flow. For the unit to operate at optimum performance the 
supply pressure from the gas cylinder would need to be lowered to 39 psig 
to give 39 psi -14 psi = 25 psid.