Agilent Technologies 22ET 8719ES Manual Do Utilizador

Página de 478
1-20
 
Making Measurements
Using Display Functions
To View the Measurement Data and Memory Trace 
The analyzer default setting shows you the current measurement data for the active 
channel.
1. To view a data trace that you have already stored to the active channel memory, press:
 
This is the only memory display mode where you can change the smoothing and gating 
of the memory trace.
2. To view both the memory trace and the current measurement data trace, press:
 
To Divide Measurement Data by the Memory Trace 
You can use this feature for ratio comparison of two traces, for example, measurements of 
gain or attenuation.
1. You must have already stored a data trace to the active channel memory, as described in 
2. Press 
 
 to divide the data by the memory.
The analyzer normalizes the data to the memory, and shows the results.
To Subtract the Memory Trace from the Measurement Data Trace 
You can use this feature for storing a measured vector error, for example, directivity. Then, 
you can later subtract it from the device measurement.
1. You must have already stored a data trace to the active channel memory, as described in 
2. Press 
 
 to subtract the memory from the measurement data.
The analyzer performs a vector subtraction on the complex data.
To Ratio Measurements in Channel 1 and 2 
You may want to use this feature when making amplifier measurements to produce a trace 
that represents gain compression. For example, with the channels uncoupled, you can 
increase the power for channel 2 while channel 1 remains unchanged. This will allow you 
to observe the gain compression on channel 2.
1. Press 
 
 to uncouple the channels.
2. Make sure that both channels must have the same number of points.
3. Press 
 
 
 to ratio channels 1 and 2, and put the 
results in the channel 2 data array. This ratio is applied to the complex data.
4. Refer to 
 for the procedure to identify the 
1 dB compression point.
Display
 MEMORY 
Display
 DATA and MEMORY 
Display
 DATA/MEM 
Display
 DATA-MEM 
Sweep Setup
 COUPLED CH OFF 
Display
 MORE   D2/D1 TO D2 ON