Agilent Technologies 22ET 8719ES Manual Do Utilizador

Página de 478
7-82
 
Operating Concepts
Limit Line Operation
Limit Line Operation
This menu can be accessed by pressing 
 
 within the system 
menu.
You can have limit lines drawn on the display to represent upper and lower limits or device 
specifications with which to compare the test device. Limits are defined in segments, 
where each segment is a portion of the stimulus span. Each limit segment has an upper 
and a lower starting limit value. Three types of segments are available: flat line, sloping 
line, and single point.
Limits can be defined independently for the four channels, up to 22 segments for each 
channel. These can be in any combination of the three limit types.
Limit testing compares the measured data with the defined limits, and provides pass or 
fail information for each measured data point. An out-of-limit test condition is indicated in 
five ways: with a 
FAIL
 message on the screen, with a beep, by changing the color of the 
failing portions of a trace, with an asterisk in tabular listings of data, and with a bit in the 
GPIB event status register B. (The analyzer also has a BNC rear panel output that 
includes this status, but is only valid for a single channel measurement.)
NOTE
The limit test output has three selectable modes. For more information, refer 
to 
Limit lines and limit testing can be used simultaneously or independently. If limit lines are 
on and limit testing is off, the limit lines are shown on the display for visual comparison 
and adjustment of the measurement trace. However, no pass/fail information is provided. 
If limit testing is on and limit lines are off, the specified limits are still valid and the 
pass/fail status is indicated even though the limit lines are not shown on the display.
Limits are entered in tabular form. Limit lines and limit testing can be either on or off 
while limits are defined. As new limits are entered, the tabular columns on the display are 
updated, and the limit lines (if on) are modified to the new definitions. The complete limit 
set can be offset in either stimulus or amplitude value.
Limits are checked only at the actual measured data points. It is possible for a device to be 
out of specification without a limit test failure indication if the point density is insufficient. 
Be sure to specify a high enough number of measurement points in the stimulus menu.
Limit lines are displayed only on Cartesian formats. In polar and Smith chart formats, 
limit testing of one value is available: the value tested depends on the marker mode and is 
the magnitude or the first value in a complex pair. The message 
NO LIMIT LINES 
DISPLAYED
 is shown on the display in polar and Smith chart formats.
The list values feature in the copy menu provides tabular listings to the display or a 
printer for every measured stimulus value. These include limit line or limit test 
information if these functions are activated. If limit testing is on, an asterisk is listed next 
to any measured value that is out of limits. If limit lines are on, and other listed data 
allows sufficient space, the upper limit and lower limit are listed, together with the margin 
by which the device data passes or fails the nearest limit.
LIMIT MENU   LIMIT LINE