7- 65
Operating Concepts
Modifying Calibration Kits
 allows you to enter the standard numbers for a TRL line or 
match calibration.
Label Class Menu 
The label class menus are used to define meaningful labels for the calibration classes. 
These then become softkey labels during a measurement calibration. Labels can be up to 
ten characters long.
Label Kit Menu 
This 
 softkey within the modify cal kit menu, accesses this menu. It is 
identical to the label class menu and the label standard menu previously described. It 
allows definition of a label up to eight characters long.
After a new calibration kit has been defined, be sure to specify a label for it. Choose a label 
that describes the connector type of the calibration devices. This label will then appear in 
the 
 softkey label in the correction menu and the 
 label in the 
select cal kit menu. It will be saved with calibration sets.
Verify Performance 
Once a measurement calibration has been generated with a user-defined calibration kit, its 
performance should be checked before making device measurements. To check the 
accuracy that can be obtained using the new calibration kit, a device with a well-defined 
frequency response (preferably unlike any of the standards used) should be measured. The 
verification device must not be one of the calibration standards: measurement of one of 
these standards is merely a measure of repeatability.
To achieve more complete verification of a particular measurement calibration, accurately 
known verification standards with a diverse magnitude and phase response should be 
used. National standard traceable or Agilent standards are recommended to achieve 
verifiable measurement accuracy.
NOTE
The published specifications for this network analyzer system include 
accuracy enhancement with compatible calibration kits. Measurement 
calibrations made with user-defined or modified calibration kits are not 
subject to the analyzer specifications, although a procedure similar to the 
system verification procedure may be used.
Saving Modified Calibration Kits to a Disk
The calibration kit, along with any calibration data and other instrument state 
information, can be saved to an ISTATE file on a floppy disk. To save a modified calibration 
kit with an instrument state, press:
 
 
 
 
 
 
TRL LINE OR MATCH 
LABEL KIT 
CAL KIT [ ] 
MODIFY [ ] 
Cal
 CAL KIT   SELECT CAL KIT   USER KIT 
Save/Recall
 SAVE STATE