A&D AD-4212B-101/102/201/301 Manual Do Utilizador

Página de 102
 
50
12.  COUNTING MODE (PC) 
This is the mode to determine the number of objects in a sample based on the standard sample unit 
mass. The unit mass is the mass of one piece of the sample. The smaller the variables in each 
sample unit mass are, the more accurate the counting will be. The AD-4212A/B series balance is 
equipped with the Automatic Counting Accuracy Improvement (ACAI) function to improve the counting 
accuracy. 
Notes 
  Use samples with a unit mass of 1 mg or more (AD-4212A-100, AD-4212B series) or 10 mg or 
more (AD-4212A-200/600/1000) for counting. 
  If the sample unit mass variable is too large, it may cause a counting error. 
  To improve the counting performance, use the ACAI function frequently or divide the samples 
into several groups and count each group. 
 
Selecting the counting mode 
1 Set 
Unit selection 
(  Unit  )
” parameter of the function table to pC (counting mode). For 
details, refer to “5-2 Changing the Unit”.
 
Storing a sample unit mass   
2  Press the SAMPLE key to enter the sample unit mass 
storing mode. 
  To return to the counting mode without storing the sample 
unit mass, press the CAL key. 
3  To select the number of samples, press the SAMPLE key 
several times. It may be set to 10, 25, 50 or 100.   
Note 
A greater number of samples will yield more accurate counting result. 
4  Place a container on the weighing pan, if necessary. Press 
the RE-ZERO key to cancel the weight (tare). The number 
specified in step 3 appears. 
 e.g.: 
 
 
25 0 pC    is displayed if 25 is selected in step 3. 
 
 
5  Place the number of samples specified on the pan. In this 
example, 25 pieces. 
6  Wait for the stabilization indicator to be displayed. Press 
the PRINT key to calculate and store the unit mass. The 
balance displays  25  pC  (counting mode) and is set to 
count samples with this unit mass. (The sample unit mass 
stored, even if the AC adapter is removed, is maintained in 
non-volatile memory.) 
  To improve the accuracy of the unit mass, proceed to step 8. 
Weighing
pan
Container
Sample
(25 pieces)