4-3.How To Operate Maintenance Mode
44    DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6
On the LUT data screen, the values (density x 100) of 16 points
of the LUT curve are displayed.
When inputting the digital value of the LUT data (12-bit digital
value converted from the density of each point: 0000 - 4095),
touch [DIGITAL] on the lower part of the screen. The digital
data of 16 points of the LUT curve will appear.
Touching [GRAPH] will allow you to confirm a graph of the
selected LUT curve.
(6) Touch [EDIT] on the LUT data screen.
The LUT data input screen (DENSITY) will appear.
(7) Touch the value of the point to be changed, and
enter a value with the numeric keys on the lower
part of the screen.
By touching [DIGITAL] / [DENSITY] on the lower part of the
screen, you can input the digital value of the density or the den-
sity value (density x 100).
(8) After the settings are complete, touch [OK].
The setting becomes valid, and the user LUT select screen re-
appears.
If you wish to cancel the settings, touch [CANCEL].
Confirming the LUT curve
Of the screens of "USER LUT" and "PRINT CONDITION",
the screen for selecting or creating the LUT can display the
graph showing the characteristics of the LUT being selected or
created by touching [GRAPH] displayed on the screen.
On the graph display screen, touching [DIGITAL] or
[DENSITY] switches between the digital value and the
density value (density x 100) for showing the vertical axis
of the graph.
Touching [EXIT] on the graph display screen will return
to the original screen.
The graph display screen only allows you to confirm the
LUT curve. You cannot edit the LUT displayed on the
screen.