Table 6-3. Performance Tests (Voltage, Resistance, and Frequency)
Function
Range
Input Level
Frequency
Display Accuracy
(1 year, 18-28
°
C)
MIN
MAX
DC Volts
300 mV
0V
--
-0.02
0.02
300 mV
150 mV
---
149.94
150.06
300 mV
290 mV
---
289.91
290.09
3V
2.9V
---
2.8991
2.9009
-3V
-2.9V
---
-2.9009
-2.8991
30V
29V
---
28.991
29.009
150V
150V
---
149.94
150.06
300V
290V
---
289.91
290.09
Note
Voltages greater than 150V can only be applied to channels 0,1, and 11.
300 mV
20 mV
1 kHz
19.71
20.29
300 mV
20 mV
100 kHz
18.50
21.50
300 mV
290 mV
1 kHz
289.28
290.72
3V
290 mV
100 kHz
275.00
305.00
3V
2.9V
1 kHz
2.8937
2.9063
30V
29V
1 kHz
28.931
29.069
150V
150V
1 kHz
149.54
150.46
300V
290V
1 kHz
289.34
290.66
Note
Voltages greater than 150V can only be applied to channels 0, 1, and 11. The rear Input Module must be
installed when measuring ac volts on channel 0.
Note
For 2-terminal measurements, the resistance accuracy given in this table applies to Channel 0 and makes
allowance for up to 0.05 ohm of lead wire resistance. You must add any additional lead wire resistance
present in your set up to the resistance values given in this table.
Resistance
Using inputs in decades of 3:
300 
Ω
short
0.00
0.09
300 
Ω
299.90
300.15
3 k
Ω
short
0.0000
0.0003
3 k
Ω
2.9989
3.0011
30 k
Ω
30 k
Ω
29.990
30.010
300 k
Ω
300 k
Ω
299.88
300.12
3 M
Ω
3 M
Ω
2.9979
3.0021