5. Troubleshooting 
REJ10J1005-0200  Rev.2.00  April 1, 2007 
 
5.2 When the Emulator Debugger Does Not Start Up Properly 
 
(1) When the LEDs of the PC7501 Do Not Display Normally 
 
Table 5.1 Errors LEDs show and their checkpoints 
Connection to 
Error 
Checkpoint 
the user system
LEDs do not light up. 
Check that the power cable is connected. 
See "2.5 Connecting the Power Supply for the Emulator" (page 23) 
and the user's manual of the PC7501. 
 
All LEDs remain lit. 
Recheck the connection between the PC7501 and this product. 
See "2.4 Connecting the PC7501" (page 22). 
 
 
The POWER LED of "STATUS 
OF TARGET" does not light up. 
Connected Check 
that 
power (Vcc) is properly supplied to the user system and 
that the user system is properly grounded. 
 
Not connected
(1) Check  that both the main and sub clocks of the emulator 
debugger are not set to "EXT". 
 
See the CLK command of the emulator debugger. 
(2) Check the oscillation circuit in the emulator is oscillating. 
 
See "2.10.2 Selecting Clock Supply" (page 40). 
 
The CLOCK LED of "STATUS 
OF TARGET" does not light up. 
Connected 
(1) When the clock is supplied from an external oscillator, check 
that the oscillator circuit in the user system is oscillating 
properly. 
(2) Check the switches in the emulation probe is properly set.. 
 
See "2.10.1 Setting Switches of Emulation Probe" (page 36). 
 
The RESET LED of "STATUS 
OF TARGET" does not go out. 
Connected 
Check that the reset pin of the user system is pulled up. 
 
 
 