DocuPen Roller Calibration Procedure
I
f your scanned image looks disproportionally stretched (or shrunk) then you need to
calibrate the rollers. 
As a rule, color calibration must be done first, before roller calibration 
1. Open the DocuPen Scanner Control window.
Click on Options and select Calibration. 
2. In the Calibration
Dialog 
window that
appears click on the Auto
button
3. In the Calibration Setup window that
appears please Calibrate Roller and click on
OK.
4. You will be prompted to turn the DocuPen ON. Press the Power On button on your
DocuPen and Click OK in the Dialog window.
Note: Your DocuPen memory will be ERASED! 
DocuPen Operation Manual 24
DocuPen Roller Calibration Procedure
After the memory has been erased the Calibration scanning window will appear. 
5. Turn the DocuPen On and scan the Roller Calibration sheet.  For optimal results
we recommend to perform the Roller Calibration in Mono Mode B/W 1 bit by Resolution
200dpi. 
6.  After the calibration image is saved into DocuPen’s memory, click OK in the
Calibration scanning window. Make sure that the DocuPen is turned ON.
If the accuracy is within 96-104%,you have successfully calibrated the roller!
7. Finally, click Upload Values button in the Calibration result window to send the
calibration result to the DocuPen. Turn ON the scanner and click OK.
DocuPen Operation Manual 25