Using Visual Programming Environments
The Agilent 816x VXIplug&play Instrument Driver
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
223
19
Press OK. The CVI Function Panel Converter window appears.
20
Press OK.
21
The Select a library window appears. Browse to 
<drive>:\vxipnp\winXX\Bin, where XX stands for NT, or 95, select 
hp816x_32.dll and click on Open.
22
The CVI Conversion Status window is displayed until the conversion is 
completed. 
You must check the Add Front Panel Controls for Size of Array Parameters 
box. There will be a front panel control created for each VI that requires 
you to assign the array size. 
N O T E
You must use the 32-bit version of the Agilent 816x VXIplug&play 
Instrument Driver with LabView 5.0.
N O T E
LabView is a trademark of National Instruments Corporation.
N O T E