Page 4 of 25                               SPEX SamplePrep LLC                                        9.24.07                        
 
 
SECTION 2.0:  6870 FREEZER/MILL SPECIFICATIONS 
 
Type of mill:  cryogenic impact mill 
Grinding mechanism:  steel impactor driven by dual electromagnets 
Coolant:  liquid nitrogen 
Weight (empty, without vial or coolant):  approx. 44 lbs. (20 Kg) 
Grinding Vials: one 6801 Large Vial, one Mid-Size Vial (6871, 6881, 6885), or one 6753 
Microvial set, also one to four Small Vials (6751, 6751/52, 6761, 6771). 
 
Typical Sample Capacity:   
Small Vials: actual volume with impactor, approx. 25 ml. Typical sample weights:  2 
grams for biological samples, 1 gram for polymers. Effective capacity for a given 
sample is determined by trials.  Overfilling a vial greatly reduces its efficiency. 
6753 Microvial Set:  100 - 500 mg per individual 6753V Microvial, depending on 
sample. 
Mid-Size Vials have up to 5 times the sample capacity of Small Vials, and Large Vials 
about twice the sample capacity of Mid-Size Vials. 
Actual performance of any vial/sample combination depends on sample properties, 
cooling time, grinding time, desired outcome, etc.  
 
Typical liquid nitrogen consumption: 
10-15 liters for initial cool-down and filling of the tub. 
4-6 liters for each hour of operation 
Actual LN consumption can vary. 
 
Electrical specifications:  CE Approved, available in 115V 60HZ or 230V 50HZ. 
Cord:  115V 60HZ version, 3-prong grounded plug supplied. 
 
230V 50HZ version, 2-prong European plug supplied. 
 
Operator is responsible for supplying alternate line cord/plug if required. 
 
Fuses:  Two 5-amp slow-blow fuses in AC Input module. 
 
25-amp slow-blow fuse on driver board 
 
Dimensions:  20½ in. (52 cm) x  21½ in. (55 cm) x 18 in. (46 cm)  
 
CE Compliance and Certification: The 6870 Freezer/Mill has been tested for compliance with CE 
Directives for Industrial, Scientific, and Medical Equipment Emission, and Generic Light Industrial 
Immunity. 
 
Safety Features:   
Lid Interlock prevents mill from running if lid is not latched shut. 
Liquid nitrogen sensor shuts down mill if liquid nitrogen level is too low. 
NOTE: Liquid nitrogen level should be visually checked during extended runs as well as 
before every run.  The Freezer/Mill warranty does not cover damage to the coil caused by 