Page 9 of 25                               SPEX SamplePrep LLC                                        9.24.07                        
 
 
SECTION 5.0: TOUCHSCREEN DISPLAYS 
 
The 6870 Freezer/Mill is programmed and operated through a series of touchscreen displays.  
Transition between screens, and all programming and operating commands, are done by touching 
the screen with a stylus.  Do not use a sharp point as this can damage or deface the screen.    
 
5.1 LOGO SCREEN 
 
When you switch on the power, the LOGO screen appears first.  This says “SPEX SamplePrep 
6870 Freezer/Mill” and includes a graphic representation of the Freezer/Mill mascot.  Touch the 
LOGO screen anywhere to transfer to the CURRENT SETTINGS  screen. 
 
5.2 CURRENT SETTINGS SCREEN 
 
This screen displays whatever program is currently set up to run.  Here is a typical program, as 
displayed: 
   
 
 
   CYCLES 
     
               
 
 PRECOOL 
 
         10 MIN 
 
 RUN TIME 
 
          2 MIN 
 
 COOL TIME 
 
         2 MIN 
 
     RATE 
 
       10 CPS 
 
   SETTINGS 
 
      Start SCRN 
 
      
          
CURRENT SETTINGS SCREEN 
 
If the Freezer/Mill has a loaded vial in the coil, is filled with LN, and the lid is down and 
latched, this is what happens when the mill is started with these settings: 
 
1)  Precool: the coil is inactive while the vial cools under LN for 10 minutes. 
2)  First cycle: the coil is activated and the impactor grinds for 2 minutes at a rate of 10 
cycles per second (20 impacts per second).  
3)  Cool time: the coil is inactive for 2 minutes while it and the sample cool back down.   
4)  Second cycle: the coil is activated and the sample is ground for another 2 minutes. 
5)  Cool time: the coil is inactive for 2 minutes.