Philips V200 Manual Do Utilizador

Página de 328
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-3
Chapter 13
Oxygen Sensor Option
The optional Oxygen sensor (O
2
 sensor) may be installed to allow monitoring of 
delivered O
2
 to the inspiratory limb of the patient circuit. The ventilator will 
accept the MSA MiniOX
®
 O
2
 and Analytical Industries Inc. sensors, P/N 
1001454.
Figure 13-1: O
2
 Sensor exploded view
Assemble O
2
 Sensor
Refer to Figure 13-1 for the following instructions.
MSA MiniOX O
2
1. Remove the O
2
 sensor (a) from the package. Discard the o-ring and 
adapter (not shown) provided with the sensor.
2. Place flat gasket (b) into the threaded hole of the O
2
 Sensor Tee (c). 
3. Screw the O
2
 sensor (b) into the O
2
 sensor tee (c).
4. Insert one end of the O
2
 sensor cord (d) into ventilator. Insert other 
end of O
2
 sensor cord (d) into the sensor. Fasten each end with 
knurled collars.
Analytical Industries Inc.
1. Remove the O
2
 sensor (a) from the package.
2. Screw the O
2
 sensor (b) into the O
2
 sensor tee (c).
3. Insert one end of the O
2
 sensor cord (d) into ventilator. Insert other 
end of O
2
 sensor cord (d) into the sensor. Fasten each end with 
knurled collars.