Philips V200 Manual Do Utilizador

Página de 328
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-17
Chapter 13
5. When using oxygen from a wall supply, attach a wall oxygen hose (not 
included) to the center inlet of the oxygen manifold assembly.
Replacement Parts:
The following replacement parts can be ordered from Respironics:
Check Valve Assembly (PN 1007190)
24-inch Medical Hose Assemblies - Oxygen (PN 1006655)
Bracket and Hardware Kit (PN 1062294)
Using the Manifold
To Use Wall Supply Oxygen
Connect the wall oxygen supply to the manifold. Close the O
2
 cylinder valves. 
Oxygen is delivered to the ventilator from all sources that are active. Therefore, 
if the O
2
 cylinder valves are open while oxygen is being delivered from the wall 
supply, the cylinders can be depleted of oxygen. To avoid depleting the O
2
 
cylinders, shut off the cylinder valves when using the wall oxygen supply.
To Use O
2
 Cylinders
Open the O
2
 cylinder valves and disconnect the wall oxygen supply. 
Transport Use
When preparing for intra-facility transport use, first open the O
2
 cylinder valves 
and then disconnect the wall oxygen supply. When transport is complete, 
reconnect the wall oxygen supply and then close the O
2
 cylinder valves.
WARNING:
If using wall oxygen supply, close the O
2
 cylinder valves to avoid 
depleting the cylinders.