Chapter 13
13-36
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Analog Output (Chart 
Recorder)
The Communications-Print option enables the ventilator to send ventilator data 
to a chart recorder or similar device using the analog output port.
Data displayed via the chart recorder is for reference purposes only. Decisions 
for patient care should not be based solely on the data displayed via the chart 
recorder.
Table 13-3 summarizes the Analog Output pinout.
X
HIGH INT 
TEMP
High 
Temperature
Temperature within ventilator enclosure exceeds 
maximum limit.
X
HIGH INT 
O2%
High Internal 
O2
Oxygen concentration within ventilator enclosure 
exceeds maximum limit.
NOTE: The following alarm messages are not displayed on the CMS:
Bad 
ADCWrap 
Sensor
ERROR
Bad Bat 
Volt 
Sensor
Service Due 
Bad Int 
O2 Sensor
Using Default Altitude 
Bad Int 
Temp 
Sensor
Using Default Compl 
Battery 
Charging
Using Default Settings
Available 
on 
CareNet 
CMS 
message
V200 
equivalent
Comments
Respironics V200 Ventilator Analog Output Port Pinout
Pin Number
Signal
1
Flow signal return
2
Volume signal return
3
Pressure signal return
4
Alarm signal return
5
Nebulizer signal return
9
Flow (0 V = -200 LPM, 5 V = 200 LPM)
10
Volume (0 V = 0 ml, 5 V = 2500 ml)
11
Pressure (0 V = -20 cmH
2
O, 5 V = 120 cmH
2
O)
12
Alarm (0 V = active alarm, 1.5 V = no alarm or silenced alarm)