MIP/P0.1 Maneuver
The MIP/P0.1 screen (Figure 13.J4) allows you to perform a Maximum 
Inspiratory Pressure (MIP) or occlusion pressure at 100 ms (P0.1) maneuver. 
It displays the maneuver results graphically and numerically. The MIP 
maneuver is available only in VCV and PCV. It is not available in NPPV.
How to perform an MIP or P0.1 maneuver
1. Enter the MIP/P0.1 screen: touch the Mechanics icon button, then 
MIP/P0.1.
2. MIP maneuver: hold down the Start MIP Press & Hold button, and 
instruct the patient to inhale several times as forcefully as possible. 
The maneuver automatically ends after 30 seconds or when you 
release the Start MIP Press & Hold button. (The Start MIP Press & 
Hold 
button can be pressed and held at any time during a breath. If 
pressed during the inhalation phase of a breath, the MIP maneuver 
will be executed during the exhalation phase of that breath. If pressed 
during an exhalation, the maneuver will be executed during the same 
exhalation.)
P0.1 maneuver: touch the Start P0.1 button. A message above the 
graphic display shows the results for the just-completed breath. The 
maneuver automatically ends after four preselected patient-initiated 
breaths have occurred or one minute has elapsed, whichever comes 
first.
3. Once touched, either one of the Start buttons changes to Stop. The 
Stop button allows you to cancel the maneuver at any time. A high-
priority alarm also cancels the maneuver.
4. At the end of a successful maneuver, a pop-up screen displays the 
result of the maneuver and asks you to accept or reject the results by 
touching the Accept or Reject button.
Information type
Description
Flow-Volume or Volume-
Time
 button
Flow-Volume: Displays a real-time flow-volume loop.
Volume-Time: Displays a real-time volume-time waveform.
Start button
Starts the maneuver. Changes to Stop when the maneuver is started. 
Changes to Continue when the maneuver is complete or when the Stop 
button is touched. Changes to Start when Continue is touched.
Vital Capacity window
Displays the most recently accepted VC maneuver results. The date 
and time of the maneuver appear below the window.
Scale (button at left of 
graphic display)
Allows you to adjust graphic scales manually. The default scales for the 
Vital Capacity screen are –1000 to +1500 mL for volume, -100 to 
+100 LPM for flow. Default scales are in effect every time you enter 
the Vital Capacity screen.
Rescale button
Allows software to adjust the vertical scales for waveforms and vertical 
and horizontal scales for loops.
Table 13-5: Summary of Vital Capacity Screen Information