Chapter 13
13-104
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Do not use with neonatal tracheostomy tubes.
Do not use on neonates.
Do not inflate the tracheostomy tube cuff until Speaking Mode has 
been turned off. 
Do not use on patients who are dependent on PEEP therapy.
Do not use with HME filters. 
Discontinue using Speaking Mode promptly and institute appropriate 
ventilation if the patient experiences difficulty or their status 
deteriorates from baseline parameters.
Perform an airway patency test immediately after activating Speaking 
Mode.
NOTE:
Patient Preparation
Using Speaking Mode on a patient requires preparation before activating the 
mode. Refer to the Sequence for Use diagram in Figure 13-58 and follow the 
necessary steps before starting Speaking Mode and discontinuing Speaking 
Mode. The Speaking Mode Clinical Guide describes patient care steps in 
detail.
WARNING:
Read and understand these Speaking Mode Option instructions before using 
Speaking Mode on a patient.
WARNING:
Wait at least 72 hours after a tracheostomy before initiating Speaking 
Mode.