Philips V200 Manual Do Utilizador

Página de 328
Appendix C
Alarm Testing Procedure
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix C-3
4. Alarm should reset. (If the backup battery is attached, the ventilator 
should return to AC power. If no battery is attached, the ventilator will 
resume operating.)
Apnea Alarm Test
1. Set the rate to 1 breath per minute.
2. Wait for 20 seconds.
3. The Respironics V200 should begin ventilating at a rate of 20 BPM 
while activating the Apnea alarm.
4. Reset the rate to 10 BPM.
5. RESET the alarm.
6. The Respironics V200 should begin ventilating normally.