Chapter 5
Setup
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
5-11
You can have the ventilator compensate the volumes delivered in volume 
controlled, mandatory breaths with the tubing compliance volume by 
activating the COMPLIANCE button (Figure 5-11). Circuit compliance is 
activated when this button has a white background. When the function is 
activated the exhaled volumes reported by the ventilator will also be tubing 
compliance compensated.
Backup Battery
You can have the ventilator confirm the backup battery is connected each time 
that the machine powers on. Pressing the BKUP BATTERY button (Figure 5-
12) all
ows this confirmation feature, which is identified by an active button 
with a white background.
Figure 5-12: Backup Battery Activated
From then on when the machine powers on, it searches for a backup battery. If 
the backup battery is connected to the ventilator, the startup is normal. If the 
backup battery is not connected, the ventilator displays a message at startup 
and a 5002 Diagnostic Code will be logged.
Figure 5-13: No Backup Battery Connected message
Backup Battery 
(confirm at 
startup) 
button active