IBM 1000-90 Manual Do Utilizador

Página de 22
COMM.EXE
7
Sending Files to the Computer
LAI-2000Plant Canopy Analyzer
The LAI-2000 is configured as a DCE device.  Connect the LAI-
2000 to your computer using the printer cable provided with the
instrument.  To send files from the LAI-2000 to a computer, use the
following functions, found in the COMM execute list:
FCT 31 - Configure the I/O port for 4800 baud, 8 data bits, no
parity, and xon/xoff = NO.
FCT 33 - Set the desired output format (Normal, with or without
observations, or Spreadsheet format).
FCT 32 - Print the desired range of files.
LI-1000 DataLogger
The LI-1000 is configured as a DTE device.  Most computers can be
connected to the LI-1000 using the 1000-03 printer cable and a
female-to-female gender changer.  To send files from the LI-1000 to
a computer, follow these steps:
1.
Press 
OUT
 on the LI-1000.
2.
Select the proper baud rate (toggle with any key except 
ENTER
).
"Baud = 3" is 300 baud, and "baud = 48" is 4800 baud.  Press
ENTER
.
3.
Select either "
V
" for vertical format or "
H
" for horizontal format.
Vertical format gives one number per line, and horizontal format
gives as many numbers per line as will fit in the line length
specified.  Press 
ENTER
.
4.
If "H" was chosen in step 3, the next display will be "len=".
Enter the maximum number of characters in a line for your
computer or printer.  Press 
ENTER
.  The following display is
"dump all=yes" (no) (LI-1000 software version 2.xx only).
Toggle yes or no, and press 
ENTER
.  The LI-1000 display will
show "dumping".