 Installation  45
 INPUT POWER SOURCE
 
 Refer to the applicable paragraphs below for information on the input power source. Do not  apply power to the power
supply until directed to do so in Chapter 3.
 
 
 Check the line
 label on the rear of your supply and verify that the voltage shown there
 corresponds to the nominal line voltage of your power source. If it does not, see "Appendix C - Line
 
Voltage Conversion" for instructions on changing the power supply’s line voltage configuration.
 
 Series 664xA and 665xA Supplies
 
 The supplied cord connects to the power receptacle on the rear panel (  2, Figure 2-l).
 
■ 
You can operate your supply from a nominal l00 V, 120 V, 220 V, 230 V, or 240 V single-phase ac power source as
indicated on the rear panel line 
 label 1.
■ 
See "AC Input Ratings" in Table l-la or Table 1-2a for the voltage and frequency range for each type of power source.
"Maximum AC Line Current Ratings" in Table l-lb or Table 1-2b show the maximum load current.
■ 
The line fuse is located in a fuseholder on the rear panel 
3
. The rear panel label 1 shows the fuse value used in the
power supply and Table 1-6 identifies the replacement fuse.
 
 
 
 Figure 2-1. Series 664xA and 665xA Power Connection
 
 
 Series 667xA Supplies
 
 
Note
This product requires single-phase input voltage.
 
 You can operate your supply from a nominal 200 V or 230 V, single-phase power source, or from the line-to-line voltage of
a 208-volt, 3-phase source. The proper source is indicated on the rear 
 label ( 4, Figure 2-2). See "AC Input
Ratings" in Table 1-3a for the voltage and frequency range for each type of power source.
 
 
Note
The power source must be a dedicated line with no other devices drawing current from it.
 
 The line fuse is located inside the power supply. Table 1-6 identifies the replacement fuse. See "In Case of Trouble" in
Chapter 3 for instructions on fuse replacement.