Environment
172
Appendix B    Safety and Environment    
The Monitor is designed to fulfill safety requirements according to IEC 60601-1, UL 544 
and CSA C22.2-601.1-M90. It is not intended to be used for direct cardiac application and 
it is not defibrillator-protected.         
The external Toco transducer (M1355A) is equipment type B 
.
M1351A: 
Connecting the transducer to the Toco channel results in B 
condition.
M1353A: 
Connecting the transducer to the Toco channel results in CF 
condition.
The IUP pressure transducer (CPJ840J5) IUP quartz transducer (1290A and the 1290C) 
are equipment type CF 
. Connecting the transducers to the Toco channel results in CF 
condition.
The ultrasound transducer (M1356A) is equipment type B 
. Connecting the 
transducer to the US channel results in B condition.
M1353A: 
The DECG transducer (M1357A) and MECG transducer 
(M1359A) are equipment type CF 
. Connecting the 
transducer to the ECG channel results in CF condition.
Environment 
Use the monitor in an environment that is reasonably free from vibration, dust, corrosive 
or explosive gases, flammable agents, extremes of temperature, humidity and so forth. It 
operates within specifications at ambient temperatures between 0 and 55
o
C. Ambient 
temperatures that exceed these limits can affect the accuracy of the monitor and cause 
damage to the components and circuits. Only products that fulfil the necessary safety and 
electrical standards should be used in conjunction with the monitor (contact your local 
response center for details). 
Allow at least 5cm (2in) clearance around the monitor for proper air circulation. If the 
monitor is mounted in a cabinet, allow sufficient space at the front for operation and at the 
rear for servicing with the cabinet door open.