â Laboratory and Chromatography Refrigerators
Page 8 of 12
Part Number: 104416  
 
Fluorescent Lamps 
 
 
 
 
 
 
Convenience Outlet 
(Chromatography 
Refrigerators Only) 
 
 
 
 
 
 
 
 
 
 
 
Manual/Automatic 
Condenser Fan 
 
 
 
 
 
The interior lamp is controlled by a rocker switch on the header panel.  
This light may be operated any time the cabinet power is turned on at 
the keyed power switch. If the interior lamp fails, replace with the  
same size and wattage lamp. DO NOT USE REDUCED WATTAGE 
LAMPS.  The reduced wattage lamps generally fail to light below 60 
°F (15 °C). 
 
 
 
Chromatography Refrigerators come equipped with a convenience 
outlet located in the center of the back wall.  Model 172G’s (three 
doors) have two convenience outlets.  The total power that can be 
supplied by the outlet (or outlets in 3-door models) is 5A @ 120VAC 
(4A @ 120VAC for the 3-door models).  The outlet(s) is protected by 
a circuit breaker located just below and to the right of the outlet(s).  
Pressing the breaker button resets the breaker after a fault. 
 
 
 
WARNING:  The convenience outlet is LIVE at any time the 
unit is connected to an electrical power source, regardless 
of the position of the key switch. 
 
 
A rocker switch located on the left side of the top of the inside of the 
chamber is used to operate the condenser fan automatically or 
manually.  With the switch in the automatic position, the condenser 
fan operates with the compressor.  In the manual position, the 
condenser fan operates continuously.   
 
During times of high ambient relative humidity conditions, more water 
is condensed on the evaporator and directed to the condensate pan.  
The condenser fan will run continuously in the manual setting to 
better dispose of this water.  At times of low relative humidity, the 
switch can be set to automatic.