CALIBRATE CTRL Submenus
143
Gap Windowing
This feature compensates for any early falling edges or spurious peaks and 
troughs that may appear within the gap length in media. These edges or 
peaks and troughs can cause unreliable detection of the leading edge of the 
next label (top-of-form). Use Gap Windowing to resolve the following 
problems:
Loss of one or more complete (serialized) labels.
Start of an image printed in the middle of a gap, especially with fanfold, 
perforated media.
Top part of an image lost when printing in head-first orientation.
The options for Gap Windowing are Disable and Enable:
Disable. When the leading edge of a gap is detected, the printer 
continuously looks for the leading edge of the next label and uses it as the 
TOF position. Perforations or unusual media discrepancies within the gap 
can cause inaccurate TOF detection.
Enable. When the leading edge of a gap is detected, the printer ignores 
the first 90% of the gap length value specified in the Gap Length menu 
option. The result is that cross perforations or unusual media 
discrepancies within the gap are filtered out, allowing the printer to 
reliably detect the actual leading edge of the next label and use it as the 
TOF position.
The factory default is Disable.
Gap Length
Gap Length is the actual length (height) of a label gap measured in .01 inch 
increments. The range is 0.05 to 1.00 inches.
NOTE: You must enter the correct Gap Length. If the Gap Length is too long, 
the image will shift down from the leading edge (TOF) of the label.
The factory default is 0.12 inches.
Cal in Peel Mode
This option allows you to perform a calibration (Auto Calibrate or Pwr Up 
Auto-Cal) in Peel-Off Media Handling mode.
Disable. The printer will not permit calibration and a
“CANNOT CALIBRATE/Disable Peel-Off” message will briefly display. 
Additionally, if “Pwr Up Auto-Cal” is enabled, the printer will not perform 
calibration at power up.
Enable. Auto Calibrate can be performed from the front panel, and if the 
Pwr Up Auto-Cal option is enabled, calibration will be performed at power 
up.
NOTE: Calibration in Peel-Off mode does not stop and wait for you to remove 
peeled labels. Therefore, be prepared to remove the labels as they 
are automatically peeled.
The factory default is Disable.