4 Basic Operations
Connecting Cables 
Connect the CR-2 AF to a computer in which the supplied software, Rics NM 2 has been installed.
 WARNING
•  Do not plug or unplug the power plug with wet hands.
•  Do not use any power cord other than the one supplied.
Otherwise, fire or electric shock may result.
Make sure that the power switch for the CR-2 AF is turned off. 
Connect the power cord and the USB cable to the 
retinal camera. 
Insert the power plug all the way into an AC outlet. 
Connect the other end of the USB cable to the computer’s USB port. 
Use a high-speed AB type USB cable that is shorter than three meters. For details, contact your 
sales representative or local Canon dealer.
Preparing for an Examination
 CAUTION
•  To prevent the risk of infection, wipe the forehead rest with disinfectant ethanol for each patient. 
For details on how to disinfect, consult a specialist.
•  To ensure cleanliness, replace the chin rest paper for each patient. 
Turn on the computer and start the Rics NM 2. 
If the digital camera is off, turn the switch to the ON. 
If the Mode Dial is set to anything other than  , an image cannot be taken.
While holding down the button at the center of the Mode Dial, turn the dial to  .
L-IE-5208A_CR-2_AF.indb   22