3-6     Operation
Rosemount Analytical Inc.    A Division of Emerson Process Management
Model 7003M
c. Error 
Messages
If an error condition occurs, E-# (non-fatal
error) or F-# (fatal error) will flash on the
display.  Table 3-10  through Table 3-13
list error codes and corrective actions.
Pressing "ACK" or "ENTER" will
acknowledge any alarm and the error
message will disappear.
CODE
DEFINITION
F-1
Invalid PROM (16#C000#..16#FFFF#) checksum detected
F-2
Internal RAM (16#00#..16#FF#) failed testing
F-3
External RAM (16#4000#..16#5FFF) failed testing
F-4
EEPROM (16#B600#.. 16#B7FF#) bulk erase failure
F-5
Signal Board not connected
F-6 thru F-9
EEPROM write failure
Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors
CODE
DEFINITION
E-1
5 volts power supply above 5.5 V
E-2
5 volts power supply below 4.5 V
E-3
12 volts power supply above 13.2 V
E-4
12 volts power supply below 10.8 V
E-5
sensor output too low for calibration
E-6
sensor missing
E-8
sample temperature is below 1
°C
E-9
sample temperature is above 44
°C
Table 3-11. RUN Mode Errors (Non-Fatal Errors
CODE
DEFINITION
HI
sample temperature is above 50
°C
LO
sample temperature is below 0
°C
Ahi
excessive sensor signal
Table 3-12. Signal Board Errors
CODE
DEFINITION
COP
Computer Operating Properly error (error detected)
iot
Invalid Op code Trap (internal computer command is invalid)
E-0
power interruption occurred
Table 3-13. Microprocessor Errors That Reset The System