Adam Equipment gbc-gfc series Manual Do Utilizador
P a g e
28
© Adam Equipment Company 2011
12.0 SERVICE
PARAMETERS
12.1 USING
“0000”
TO
ENTER
THE
CALIBRATION
PARAMETER
•
Press the [Tare] key during the display counting when turned on. The
display will ask for a code number, “Pin “ on the Weight Display.
display will ask for a code number, “Pin “ on the Weight Display.
•
Enter the number 0000 when “Pin “ is displayed and then press
[Tare]. If another PIN number has been previously stored enter that
number instead.
[Tare]. If another PIN number has been previously stored enter that
number instead.
•
The displays will show the first parameters, called “F1” “CAL”.
•
To select another parameter press the [U.Wt./Units] key to advance
through the parameters.
through the parameters.
•
Press the [Tare] key to enter a parameter.
•
To exit a parameter, press the [Zero] key.
•
The Weight window will show the parameter number and the Unit
Price and Count windows will show the word describing the function.
Price and Count windows will show the word describing the function.
•
When a parameter is entered by pressing the [Tare] key, the displays
will guide you through the parameter selected and the options
available.
will guide you through the parameter selected and the options
available.
The parameters available are:
“F1”
“
CAL”
To enter the Calibration
“F2” “Pin”
To set a new Pin number for access
“F3” “Unit”
Enable Weighing Units