Adam Equipment abk 8 Manual Do Utilizador

Página de 48
20
 | 
P a g e                  
© Adam Equipment Company 2011
 
comparison  to  the  capacity  of  the  system.    A  smaller  weight  will  show 
only “
100%
” while a larger weight might show “
100.00%
”.    
 
  If the scale was showing zero weight when entering this function, then 
the  user  must  manually  enter  the  weight  to  be  set  as  100%  as 
described below. 
 
  Press the [Z/T] key. “
P1  PCt
” will be displayed.  
 
  Press  [Z/T] again to enter percent weighing. The scale will now accept 
a value the user enters as the reference weight.  
 
  Enter  the  value  using  the  [Unit/
]    key  and  [Pcs/ ]  key  to  set  the 
values of the 100% mass.  
 
  When set press the [Z/T] key to store the value.   The display will show 
0.00   %
”.  
 
  The weight entered must be greater than 50 scale divisions.   
 
  Press [Unit/
]  to return to weighing. 
 
NOTE:  
 
The  display  may  jump  by  large  numbers  unexpectedly  if  small  weights  are 
used to set as 100% reference.   The scale checks if the weight is too small 
and will show Error 7.