12.0 TECHNICAL PARAMETERS 
 
The technical parameters accessed via the “
tECH
” prompt at the end of user 
parameters  and  are  password  controlled  to  prevent  unauthorised  access.  
These  parameters  set  the  metrology  for  the  scales.      Each  scale  is  set 
independently.  The  parameters  will  set  capacity,  division,  decimal  point 
position,  initial  zero  range,  auto  and  manual  zero  range  as  well  as  factory 
calibration.   
 
ACTION 
DISPLAYS 
When “
tEch
” is displayed, press the 
[Print] key. The scale will ask you 
to enter the password. 
 
 
“ Pi n” 
Default Password is 0000. 
Enter “0” four times. 
2006 will override any other user 
password. Press [Print]. 
 
 
“ Pi n”     “ ----” 
Select the scale to be set up by 
using the [Local/Rem] key. 
 
Press the [Print] to enter the 
technical section for that scale.  
 
“t ECH”  “   Lo c A L”   “      “   
“t ECH”  “r E mo t E”   “      “   
 
Use the [U.Wt./Units] to select the 
weighing unit to be used for setting 
up the parameters for the scale. 
The arrow in the “Weight” window 
will indicate the unit selected. 
Press the [Print] key to continue. 
 
  
“t ECH”  “   Uni t”   “      “   
 
 
 
Press the [U.Wt./Units] key to 
scroll through the menus for the 
chosen scale. User  [Print] to 
select and set parameters and the  
[Tare/Zero] to escape as without 
changing data as with the user 
parameter section.  
 
“Cn t” 
   
Both scales have the same menus 
except that the capacity of the local 
scale  is  fixed  and  cannot  be 
entered.